ReNew Technical Reference
42
Troubleshooting Guide
Crosshead will not stop • Servo is improperly connected.
• Improper machine parameters. Contact an MTS Applications Engineer for
further assistance.
• Controller error. Contact an MTS Applications Engineer for further
assistance.
Limit is always active —
cannot clear
• Proximity limit switch on the crosshead is obstructed.
• Limit switch is defective. Contact an MTS Applications Engineer for further
assistance.
• Frame controller limit switch circuitry is malfunctioning. Contact an MTS
Applications Engineer for further assistance.
Limit switches not
functioning
• Defective limit switch. Same as above. DANGER — DO NOT RUN
MACHINE! Contact an MTS Applications Engineer for further assistance.
Will not pull rated
maximum load
• Defective servocontroller. Contact an MTS Applications Engineer for
further assistance.
• Load cell is not calibrated.
Improper machine
parameters
• Contact an MTS Applications Engineer for further assistance.
Load cell will not
calibrate
• Load cell is not connected.
• Defective load cell.
• You are attempting to automatically calibrate a non-MTS load cell. These
must be manually calibrated.
• Improper calibration information being used. Check the load cell
information in the testing software.
Load cell amplifier
defective
• Contact an MTS Applications Engineer for further assistance.
Load plateaus • Defective load cell.
Unusual noise • Grinding sound indicating that a moving part is rubbing. Lubrication is
needed. Inspect belt and screw bearings. Contact an MTS Applications
Engineer for further assistance.
• Debris caught in the cooling fan.