Mts SWIFT 10 ATV Sensor Manuale Utente Pagina 46

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 45
SWIFT 10 ATV Sensors
46
TI2SHUNT - Transducer Interface Shunt
Software Utilities
Option 2 Use this to set the shunt tolerance. When selected, the following is displayed:
Option 3 Use this option to apply a shunt to each bridge individually, read the output of the
bridge, compare the result with the value stored in the Transducer Interface and
display the difference each bridge. See the table on the next page. The shaded
fields are the bridge being shunted. The actual sequence depends on the
transducer wiring. Note that this option does not update the shunt measured
values or error status.
Enter choice: 2
The current shunt tolerance is 2%
Enter new shunt tolerance in percent?
Vedere la pagina 45