Mts SWIFT 20 Ultra Sensor Manuale Utente Pagina 58

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 57
SWIFT 20 Sensors
58
Shunt Error Status
Software Utilities
Shunt Error Status
This is the error status for each bridge from the last shunt cal that was performed.
The error status is a bit-mapped status word, where the error is present if the bit is
set to 1.
Bit 0 = Bridge amplifier was saturated prior to applying the shunt.
Bit 1 = Bridge amplifier was saturated after applying the shunt.
Bit 2 = Measured delta value failed high.
Bit 3 = Measured delta value failed low.
Bit 4 = A procedural error occurred (unable to read a value, etc.).
In the following example the error status received is 5 indicating that bits 0 and 2
were set.
Bit 43210
Bit Value168421
Example00101
Error Status=
5
Vedere la pagina 57
1 2 ... 53 54 55 56 57 58 59 60 61 62 63 ... 175 176

Commenti su questo manuale

Nessun commento