Mts TestLine Manuale Utente Pagina 102

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 132
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 101
For this example choose limits from -0.6 kN to 1.0 kN.
a) If necessary, click the Station Setup button.
b) Select Channels > Ch1 > Force > Channel Input Signals button > Limits tab.
c) Set the Upper Limit to 1.0 kN.
d) Set the Upper Action to Interlock.
e) Set the Lower Limit to -0.6 kN.
f) Set the Lower Action to Interlock.
Minimize Station Manager
The Station Manager application must be running underneath the MPX application in order to run tests.
Minimizing the Station Manager application clears the desktop while leaving it running for the MPX
application.
1. Turn Exclusive Control off.
102 | MTS TestLine Test System Operation
Running the Example HCF Test
Vedere la pagina 101
1 2 ... 97 98 99 100 101 102 103 104 105 106 107 ... 131 132

Commenti su questo manuale

Nessun commento