Controller Connections
MTS Insight™ Material Testing System Installation
47
J2 Handset This is intended to interface to the handset. Specifics for this connector are:
• 12 V output power with 200 mA current limit
• RS422 driver (differential)
• RS422 receiver (differential)
• Interlock input. Handset shorts between INTLK+ and INTLK- when it is
connected.
• 8-pin RJ-45 connector
Pin Signal
1Transmit +
2Transmit -
3+12V
4INTLK+
5INTLK-
6 Analog GND
7 Receive +
8 Receive -